Harmonized Terminology Database

Editor's Note: Standard, internationally preferred terms and definitions are highlighted in blue. Terms and definitions in black are acceptable, but only if the standard internationally preferred terms and definitions are unacceptable for a certain context. Terms highlighted in red are not acceptable in the international standards community. They appear for educational purposes only. Notes and examples included with definitions are illustrative, and are not to be considered part of the standard definition.

Terms often include related terms and/or alternate terms. Related terms may have similar meanings or apply in similar contexts as the main term, and are included to help users choose the best term for their needs. Alternate terms are synonymous with the main term. In the Definitions sections of CLSI documents, related terms are indicated with a single slash (/) between the two terms, and alternate terms are indicated with the use of two slashes (//) between the two terms.

Back to Search Form


analytical measuring interval

set of values of quantities of the same kind that can be measured by a given measuring instrument or measuring system with specified instrumental measurement uncertainty, under defined conditions (JCGM 200:2012)

NOTE 1: In some fields, the term is “analytical measurement range,” “measuring range,” or “measurement range” (modified from JCGM 200:2012); NOTE 2: The lower limit of a measuring interval should not be confused with “detection limit” (JCGM 200:2012); NOTE 3: This represents the interval of in vitro diagnostic examination results over which the performance characteristics of the in vitro diagnostic medical device were validated by the manufacturer; NOTE 4: Formerly, the terms “reportable range” or “measuring range” were used in CLSI documents; NOTE 5: The measuring interval over which the performance characteristics of an in vitro diagnostic medical device have been validated has been called the reportable range (ISO 18113-1); NOTE 6: For a discussion of the difference between interval and range, see A.2.11 (ISO 18113-1); NOTE 7: In the United States, the term often used is “analytical measuring range”; NOTE 8: In laboratory medicine, the term is defined as “measuring range,” or “analytical measurement range”; NOTE 9: The measuring interval is intended to refer to the interval of values that a measurement procedure can directly measure on the sample without any dilution, concentration, or other pretreatment not part of the usual measurement process; NOTE 10: The measuring interval has been called the analytical measurement range, reportable range, analytical range, and calibration range; NOTE 11: For EP05, the range of values (in units appropriate for the measurand) over which the acceptability criteria for the measurement procedure have been met; ie, the intersection of the linearity interval, the interval limited by the lower and upper limits of quantitation, and the interval representing acceptable imprecision; NOTE 12: The interval (or range) of values (in units appropriate for the analyte [measurand]) over which the acceptability criteria for the measurement procedure have been met; that is, where errors due to nonlinearity, imprecision, or other sources are within defined limits; NOTE 13: Formerly, the term “reportable range” was used in EP15, and another commonly used term is “analytical measurement range”; NOTE 14: The concentration range of results for which the measurement procedure functions to meet its intended use; NOTE 15: Measuring interval is determined by linearity, accuracy, and limit of detection; NOTE 16: The analytical measuring interval is the concentration value interval within which a given accuracy is achieved and a linear relationship between measurand values and the measurement procedure’s results is maintained with no preexamination dilution, concentration, or other pretreatment not part of the standard or routine measurement; NOTE 17: Although the terms “analytical measuring range” and “working interval” are commonly used instead of “analytical measuring interval,” their use is not encouraged; NOTE 18: This interval is separate from the extended measuring interval, within which preexamination dilutions are needed to obtain results; NOTE 19: It is sometimes called “analytical measurement range” or “measurement range,” which is the range of analyte values that a method can directly measure on the specimen without any dilution, concentration, or other pretreatment not part of the usual assay process; NOTE 20: The ability of a measurement procedure to detect small quantities of the analyte; NOTE 21: Analytical sensitivity is usually expressed as the “minimum detectable concentration” or “detection limit.”

Alternate Term: working interval; measuring range; measuring interval; measurement interval

Project: NBS11, I/LA21, ISO 18113-1, H26, POCT07, MM06, EP06, C51, EP17, C56, MM14, EP09, EP26, H60, VET04, C62, EP05, EP15, C57, MM03, MM23, POCT06, EP21, EP46, C24, EP34, C63, MM17, C34, C62, GP26, H26, EP39, EP35, NBS10

Related Term(s):
analytical measurement range
range of analyte values that a method can directly measure on the specimen without any dilution, concentration, or other pretreatment, not part of the usual assay procedure


measuring interval

(of an IVD medical device) set of values of measurands for which the uncertainty of an IVD medical device is intended to lie within specified limits

NOTE 1: This represents the interval of IVD examination results over which the performance characteristics of the IVD medical device have been validated by the manufacturer; NOTE 2: May be referred to as "reportable range" in some laboratory quality systems.

Project: ISO CD 18113-1, ISO CD 18113-2, ISO CD 18113-3


measuring interval

set of values of quantities of the same kind that can be measured by a given measuring instrument or measuring system with specified instrumental measurement uncertainty, under defined conditions (JCGM 200:2012)

NOTE 1: The measuring interval is intended to refer to the interval of values that a measurement procedure can directly measure on the sample without any dilution, concentration, or other pretreatment not part of the usual measurement process; NOTE 2: The measuring interval has been called the analytical measurement range, analytical range, and calibration range; NOTE 3: The concentration range of results for which the test method functions to meet its intended use; NOTE 4: Measuring interval is determined by linearity, accuracy, and limit of detection; NOTE 5: This represents the interval of in vitro diagnostic examination results over which the performance characteristics of the in vitro diagnostic medical device were validated by the manufacturer; NOTE 6: The measuring interval over which the performance characteristics of an in vitro diagnostic medical device have been validated has been called the reportable range; NOTE 7: See reportable range.

Project: ISO 17593, MM01, MM22, EP19, EP10


measuring interval

the upper and lower values, between which the test system will give accurate results

NOTE: Alternative terms are “reportable range” and “analytical measurement range.”

Project: POCT08


measuring range

See measuring interval

Alternate Term: measuring interval

Project: C56


working interval

See measuring interval

Alternate Term: measuring interval

Project: C51, VET04



© 2025 Clinical and Laboratory Standards Institute. All Rights Reserved.